Digital thermometer SIMPLE

The DIGITAL THERMOMETERS are intended to measure the human body temperature

Fast measurement in regular mode:

  • Orally
  • Under the arm
  • Rectally

UNIVERSAL USE

Can be used to measure temperature: orally, under the arm, rectall.

FAHRENHEIT/ CELSIUS SWITCH

Switch between Fahrenheit °C and Celsius °F (default is Celsius °C);

EASY TO MANAGE

Measurement is made by means of one button

CONVENIENT AND PRECISE

Measure Range: 32,0℃ - 42,9℃ (90,0℉ - 109,9℉);
Accuracy: ±0.1℃ (±0.2℉) during 35.5℃ ~ 42.0℃ (95.9℉ ~ 107.6℉)

IDEAL FOR TRAVEL

Stylish travel case - keep the thermometer safe;
The device's idle time is 10 minutes, after which the power turns off automatically;

FEATURES

Measure Range: 32.0℃-42.9℃(90.0℉-109.9℉ )( ℃ /℉ chosen by manufacturer)

Accuracy: ±0.1℃ (±0.2℉) during 35.5℃ ~ 42.0℃ (95.9℉ ~ 107.6℉) at
 18℃ ~ 28℃ (64.4℉ ~ 82.4℉ ) ambient operating range; ±0.2℃(±0.4℉) for other measuring and ambient operating range

Display: Liquid crystal display

Memory: For storing the last measured value

Battery: One 1.5 V DC button battery (size LR41 or SR41, UCC 392)

Battery life: Approx. 200 hours of continuous operation or 1 year with 3 measurements per day

Dimension: 13.8 cm x 2.2 cm x 1.2 cm (L x W x H)

Weight: Approx. 12 grams including battery

Expected service life: Three years

Ambient operating range:

Temperature: 5℃~40℃ (41℉~104℉)

Relative humidity: 15% ~ 95% RH

Atmospheric Pressure: 700 hPa ~ 1060 hPa

Storage and transportation condition:

Temperature: -20℃ ~ 55℃(-4℉~131℉)

Relative humidity: 15% ~ 95% RH

Atmospheric Pressure: 700hPa ~ 1060hPa

Ingress Protection Rating: IP 27

Classification: Type BF

  • 1 Thermometer
  • 1 Owner’s Manual
  • 1 Storage Case

How to get more accurate measurement results?

The minimum measurement time until the signaling tone must be maintained without exception. The measurement continues even after the buzzer notification. So that in order to achieve better body temperature measurement result, recommend to keep the probe in mouth and rectum about 2 minutes, or in armpit about 5 minutes regardless of the beep sound and at least 30 seconds measurement interval should be maintained.

How to replace the battery yourself?

Pull battery cover off; gently pull out plastic circuit board with battery chamber approximately 1 cm; Use pointed object such as a pen to remove old battery. Discard battery lawfully. Replace with new 1.5V DC button type LR41 or SR41,UCC392, or equivalent. Be sure battery is installed with polarity facing up; Slide battery chamber back into place and attach cover;

Can this thermometer be used to measure the temperature in the ear canal (in the ear)?

Do not use thermometer in ear. Designed use is for oral, rectal, and armpit (axilla) readings only.

How to switch the measurement scale between degrees ℃/℉?

Temperature readings are available in Celsius (℃) or Fahrenheit (℉) and are displayed in the top right corner of the LCD. With the unit off, press and hold the On/Off Button for approximately 2 seconds to change the current setting.